DxH 900 Hematology Analyzer
$0.00$0.00
The DxH 900 hematology analyzer delivers accurate and high-confidence results on the first run. With its best-in-class hematology technology, this automated analyzer minimizes slide reviews, maximizes laboratory staff time, and provides significant cost savings. The DxH 900 offers scalable, customized solutions to address the hematology testing needs of high-volume labs.
Description
Technical Specifications
Methodology |
Enhanced Coulter Principle, VCS 360 with DataFusion | |||||
ParametersEmpty heading |
CBC:Â WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV
Differential: NE, LY, MO, EO, BA, NRBC, NE#, LY#, MO#, EO#, BA#, NRBC #, MDW (when enabled) Retic: RET, RET#, MRV, IRF Body Fluids (Cerebrospinal, Serous, or Synovial): RBC, TNC |
|||||
Throughput (tests per hour) |
Up to 100 samples/hour | |||||
Sample Volumes |
165 μL aspiration volume; 0.5 ml dead volume. Predilute: 50 μL whole blood into 200 μL of diluent | |||||
Analytical Measuring Ranges2 |
Whole Blood (CBC) | |||||
| Parameters | Units | Analytical Measuring Range | ||||
| WBC | x103 cells/μL | 0.050–400.000 | ||||
| RBC | x106 cells/μL | 0.005–8.500 | ||||
| HGB | g/dL | 0.10–25.50 | ||||
| MCV | fL | 50.00–150.00 | ||||
| RDW | % | 10.00–40.00 | ||||
| RDW-SD | fL | 15.00–150.00 | ||||
| PLT | x103 cells/μL | 3.0–3000.0 | ||||
| MPV | fL | 5.00–25.00 | ||||
| Body Fluids | ||||||
| TNC | cells/mm3 | 20–89,000 | ||||
| RBC | cells/mm3 | 1,000–6,200,000 | ||||
Data Management |
Up to 50,000 patient results with histograms, scatter plots, and demographics
100,000 patient results with graphics for power computer |
|||||
Quality Assurance |
QC with Levy-Jennings graph; XB/XM for moving averages, daily check, intelligent quality monitoring, MRV, IRF, customizable calibration and QC reminders and alerts, auto export of QC | |||||
Acoustic Noise Level |
≤ 60 dBa | |||||
Radio Frequency and Interference Immunity |
Meet the product level standards EN 61326-1 (Electrical Equipment for Measurement, Control, and Laboratory Use—EMC Requirements), and EN 61326-2-6 (Part 2, Specific Requirements for In-Vitro Diagnostic Devices) | |||||
Power Consumption |
DxH 900 SPM: 520 W (1775 BTU/hour), Monitor: 35 W (120 BTU/hour), DxH Power Computer: 160 W (546 BTU/hour) | |||||
Dimensions and weight3 |
Width | Height | Depth | Weight | ||
| 76.2 cm (30 in.) | 176.53 cm (69.5 in.), with cover closed
199.39 cm (78.5 in.), with cover open |
83.82 cm (33 in.), including the removable back panel
80 cm (31.5 in.), excluding the removable back panel |
~292.11 kg (644 lbs.), including monitor & computer ~287.12 kg (633 lbs.), |
|||
Additional Clearance |
15.2 cm (6.0 in.) per side of instrument for troubleshooting; 3.8 cm (1.5 in.) behind instrument for sufficient cooling | |||||
Maintenance and Troubleshooting
Service |
New Preventative Maintenance (PM) schedule with customer care visits |
Support |
Onboard IFU and Instructional Videos, Automated QC and calibration reminders, PROService |
Analyzer Downtime |
Minimum of 30 minutes in cleaner every 24 hours with automatic daily checks |