DxC 500i Clinical Analyzer

$0.00$0.00

Where Flexibility Drives Productivity

Meet your partner for integrated testing with a simplified workflow. The space-saving DxC 500i Clinical Analyzer delivers high-quality chemistry and immunoassay results you can trust for your low-volume lab.

Description

Main Specifications

Analytical method

Fully automated, random-access integrated system with clinical chemistry and immunoassay modules

Analytical principles

Spectrophotometry, potentiometry, and chemiluminescence

Assay types

  • Chemistry: Endpoint, rate, fixed point, indirect ISE
  • Immunoassay: Competitive assay, Sandwich assay, and Antibody detection assay

Analytical method

  • Chemistry: Colorimetry, turbidimetry, latex agglutination, and indirect ISE
  • Immunoassay: Paramagnetic coated antibody/antigen

Chemistry menu capacity

>170 preprogrammed assays currently available

Throughput

  • 400 photometric tests per hour, up to 800 with ISE
  • 100 immunoassay tests per hour (one-step assays)
  • ISE maximum test per hour: 600 if ISE only
  • HbA1c maximum tests per hour: 65 if HbA1c only

Sample types (assay dependent)

Serum, plasma, urine, CSF, whole blood, and amniotic fluid

Sample management/capacity

  • 168 sample tubes (24 racks of 7 tubes), unique continuous loading feature
  • Independent analysis:
    • Chemistry: 22 samples in refrigerated STAT carousel
    • Immunoassay: 60 samples in 6 carousel racks

Barcoded reagents

Automatic tracking of: available tests, expiration date, lot number

Sample container sizes

  • Sample tubes:
    • 13 x 75 mm, 13 x 100 mm, 16 x 100 mm
    • Sarstedt S-Monovette 4.9 mL, 5.5 mL, 7.5 mL
  • Cups/insert cups:
    • 0.5 mL, 2.5 mL, 3.0 mL cups
    • 1.0 mL, 2.0 mL insert cups, pediatric insert
    • Auto aliquot

Sample volume (assay dependent)

  • Chemistry: 1.0 –25 μL in 0.1 μL increments (assay dependent)
  • Immunoassay: 10–110 μL

Sample barcode symbology

Codabar (NW-7), CODE 39, CODE 128, ISBT-128, Interleaved 2 of 5

Compartment temperatures

  • Chemistry:
    • Chemistry reagent compartment: 39.2°F to 53.6°F (4°C to 12°C)
    • STAT wheel: 39.2°F to 53.6°F (4°C to 12°C)
    • Cuvette wheel: 98.6°F ± 0.5°F (37°C ± 0.3°C)
  • Immunoassay:
    • Reagent compartment: 33°F to 41°F (1°C to 5°C)
    • Maintains air and reagent temperature at 37.4°F to 50°F (3°C to 10°C)
    • Incubator and wash/read wheel: 97.5°F ± 0.5°F (36.4°C ± 0.3°C)

Reagent storage capacity

  • Chemistry: 76 positions in a refrigerated reagent carousel that holds 15 mL, 30 mL, and 60 mL bottle sizes
  • Immunoassay: 24 positions in refrigerated reagent carousel

Reaction cuvette

  • Chemistry: Permanent glass cuvettes
  • Immunoassay: Disposable reaction vessels

Sample quality analysis

  • Lipemia, hemolysis, and icterus indices
  • Clot detection and probe crash protection

Calibration (assay dependent)

  • Chemistry: calibration curve stability up to 90 days
  • Immunoassay: calibration stability up to 63 days
  • Calibration curves and parameters displayed on screen
  • Calibrator setpoint entry through barcode scan

Quality control

  • Shewhart day-to-day management (Levey-Jennings method)
  • Multi-rule control (Westgard method)

Installation Requirements

Dimensions and weight

  • Analyzer with monitor:
    • Weight: Dry, 1,430 pounds (<650 kg)
    • Length: 100 in (254 cm)
    • Height: 48.8 in (124 cm)
    • Height with monitor: 63.0 in (160 cm)
    • Depth: 30.3 in (77 cm)

Clearance Requirements

  • Clearance around the analyzer for ventilation:
    • Rear: minimum of 24 in (64 cm)
    • Left: minimum of 20 in (50 cm)
    • Right: minimum of 20 in (50 cm)
    • Front: minimum of 25 in (63.5 cm)>
    • Top (from the top of the chemistry analyzer): Minimum of 49 in (124 cm)

Waste requirements

  • Chemistry module drain location:
    • Maximum distance: 33 ft (10 m) from the system
    • Maximum height: 5 ft (1.5 m) from the floor
  • Chemistry module drain volume:
    • Concentrated waste: 5.5 L/hour (normal), 7.5 L/hour HbA1c
    • Diluted waste: 12 L/hour
  • Immunoassay module drain location:
    • Maximum distance 9.8 ft (3 m) from the system
    • Maximum height 3 ft (0.9 m) from the floor
  • Immunoassay module drain volume:
    • 1 L/hour
  • Immunoassay module solid waste: (RVs):
    • RVs collected in RV waste bag (300 RV capacity)

Water supply requirements

  • Maximum water consumption: 20 L/hour
  • Water type: Deionized CAP type II or better, bacteria free
  • Continuous flow supply
  • Resistivity: > 0.5 MΩ∙cm filtered with a 0.5 μm filter

Power and environmental requirements

Power supply: 200V, 208V, 220V, 230V, 240V, 50/60Hz, 3.5kVA

Computer

  • Storage: 960GB SSD
  • Operating system: Windows 10
  • External communication types: RS232 serial interface
  • Ethernet: LAN using TCP/IP

Monitor

  • Touchscreen display: 21.5 in (54.6 cm) diagonal TFT LCD with projected capacitive touch
  • Pixels: 1920 x 1080 RGB

Operating environment

  • Ambient temperature: 64°F to 82°F (18°C to 28°C)
  • Humidity: 20% to 80% RH (without condensation)
  • Altitude: Maximum 2,000 m (6,561 ft) above sea level
  • Noise level: <75 dBA

Communication modes

Serial or TCP (1 or 2 physical) connection to LIS per ASTM protocol

Reviews

There are no reviews yet.

Be the first to review “DxC 500i Clinical Analyzer”

Your email address will not be published. Required fields are marked *

Select the fields to be shown. Others will be hidden. Drag and drop to rearrange the order.
  • Image
  • SKU
  • Rating
  • Price
  • Stock
  • Availability
  • Add to cart
  • Description
  • Content
  • Weight
  • Dimensions
  • Additional information
Click outside to hide the comparison bar
Compare