DxH 900 Workcell Automated Hematology Solution
$0.00$0.00
Solutions to Redefine Your Hematology Workflow
Streamline processes with our automated hematology system. Fully scalable and customizable, the workcell features an integrated line with a seamless connection to attached units for intelligently distribute workloads.
Description
DxH 900 Workcell Specification
| Workcell | Configuration Detail | Maximum Throughput | Width | Depth | Height | Weight** |
| DxH SMS II | 1 DxH SMS | 140 Smears/Hour | 94 cm (37 in) | 78.74 cm (31 in) | SMS; 204.5 cm (80.5 in) | 353.84 kg (780 lbs) |
| DxH 900 | 1 DxH 900 | 100 Samples/Hour | 75.57 cm (29.75 in) | 82.80 cm (32.6 in) including the removable back panel 79.25 cm (31.2 in) excluding the removable back panel | DxH 900 174 cm (68.5 in) 194 cm (76.4 in) with cover lifted | 254 kg (560 lbs) |
| DxH 900-S | 1 DxH 900 with SMS | 100 Samples/Hour | 170.53 cm (67.13 in) | 82.80 cm (32.6 in) including the removable back panel 79.25 cm (31.2 in) excluding the removable back panel | DxH 900 174 cm (68.5 in) 194 cm (76.4 in) with cover lifted | 608 kg (1340 lbs) |
| DxH 900-2 | 2 DxH 900 | 200 Samples/Hour | 152.1 cm (59.88 in) | 82.80 cm (32.6 in) including the removable back panel 79.25 cm (31.2 in) excluding the removable back panel | DxH 900 174 cm (68.5 in) 194 cm (76.4 in) with cover lifted | 508 kg (1120 lbs) excluding reagents 680 kg (1500 lbs) including reagents |
| DxH 900-3 | 3 DxH 900 | 300 Samples/Hour | 228.63 cm (90.01 in) | 82.80 cm (32.6 in) including the removable back panel 79.25 cm (31.2 in) excluding the removable back panel | DxH 900 174 cm (68.5 in) 194 cm (76.4 in) with cover lifted | 762 kg (1680 lbs) |
| DxH 900-3S | 3 DxH 900 with SMS | 300 Samples/Hour | 323.59 cm (127.39 in) | 82.80 cm (32.6 in) including the removable back panel 79.25 cm (31.2 in) excluding the removable back panel | DxH 900 174 cm (68.5 in) 194 cm (76.4 in) with cover lifted | 1116 kg (2460 lbs) |
Additional clearance; 3.8 cm (1.5 in) behind instrument for sufficient cooling, 15.2 cm (6.0 in) per side.
| Whole Blood | Reportable Range | Precision |
| WBC x103 cells/μL | 0.050–400.000 | <3.0% -5% CV |
| RBC x106 cells/μL | 0.005–8.500 | <1.5% CV |
| HGB g/dL | 0.10–25.50 | <1.5% CV |
| MCV fL | 50.00–150.00 | <1.0% CV |
| PLT x103 cells/μL | 3.0–3000.00 | <3.5% CV |
| MPV fL | 5.00–25.00 | <2.5% CV @PLT >100×103 cells/μL |
| Body Fluid | Reportable Range | Precision |
| TNC cells/mm3 | 20–89,000 | <10.0% CV |
| RBC cells/mm3 | 1,000–6,200,000 | <15.0% CV |
Sample Aspiration |
Whole Blood and Body Fluid; 165 μL aspiration volume; 0.5 ml dead volume, Pre-dilute with X5 dilution mode; 50 μL whole blood into 200 μL of Diluent, SMS; 90 μL for up to 4 smears |
|
Methodology |
Enhanced Coulter Digital Impedance, 360 VCS with DataFusion | |
Available Test Menu |
CBC: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV Differential: NE, LY, MO, EO, BA, MDW, NRBC, NE#, LY#, MO#, EO#, BA#, NRBC# Monocyte Distribution Width (MDW) (when enabled) Retic: RET, RET#, MRV, IRF Body fluids (CSF, Serous and Synovial): TNC, RBC, @BFM%, @BFP% |
|
Reagents (CBC/Diff/Retic)* and Controls |
Coulter DxH Diluent…………….628017 10 L Coulter DxH Cell Lyse…………628019 5 L Coulter DxH Diff Pack…………628020 EA Coulter DxH Retic Pack………628021 EA |
6C Plus Cell Control 12×3.5 mL………………………PN C07297 6C Cell Control 9×3.5 mL………………………………..PN A59925 6C Cell Control 12×3.5 mL……………………………….PN 628027 Latron Control 8×4 mL, CP-X…………………………PN 628024 S-CAL Calibrator 1×3.3 mL………………………………PN 628026 Reticulocyte Controls Retic-X (12×3.5 mL)…PN 628028 Body Fluid Control 12×3.5 mL………………………..PN 628030 Linearity Product Lin-X……………………………………PN 628029 |
Cleaner |
Coulter DxH Cleaner 628023 10 L—Preloaded for autonomous Daily Cleaning | |
Power Consumption |
Dual indentical power computers in a workcell. DxH 900 SPM; 520 W (1775 BTU/Hour), Monitor: 35W (120 BTU/Hour), DxH Server Computer: 160 W (546 BTU/Hour), DxH Slidemaker Stainer II SPM, 550 W (1877 BTU/Hour) | |
Data Management |
Up to 100,000 patient results with histograms, scatterplots and demographics | |
Quality Assurance |
QC with Levy-Jennings graph; XB/XM for moving averages, daily check, intelligent quality monitoring, customizable calibration and QC reminders and alerts, auto export of QC | |
Sample Transport |
Capacity: 20 five-tube cassettes per module (up to maximum of 400 samples with a DxH 900-3S). Barcodes: digital barcode with 2D barcode capability, Sample ID: up to 22 characters | |
Remote Monitoring |
PROService remote management system: maximum uptime with walk-away troubleshooting and preemptive alerts of instrument issues with RAP box | |
Power Supply Specifications |
A three-wire AC outlet, single phase power supplying 90-264 Vac at 48 to 63 Hz | |
Acoustic Noise Level |
60 dBa | |
Radio Frequency and Interference Immunity |
Meets by the product level standards EN 61326-1 (Electrical Equipment for Measurement, Control, and Laboratory Use – EMC Requirements), and EN 61326-2-6 | |